Question/Answer

what does desalting resin contain?

: Each different cartridge contains a different material to remove the analytes of interest. This particular one contains silver nitrate to remove the chloride from solution.

Do you have agent in Saudi Arabia? will you please share details of this product?

This cleanup cartridge is designed to remove chloride and other halogens from an aqueous sample. The resin is packed with silver ions to precipitate out AgX2 as the sample flows through. The cartridge should be able to remove at least 2.5 mg chloride in the sample. Because it uses silver as the complexing agent other ions such as cyanide or phosphate could be removed from solution as well. Anything that forms an insoluble silver salt has the potential to be removed.