Item No.: IH4LB - Hydrogen Sulfide Detector tube 0.5-12 ppm, 10p
Qty
Price
1
$72.00
Quantity (EACH):
Order
Performance
Measuring Range
0.5 to 1 ppm
1 to 6 ppm
6 to 12 ppm
Number of Pump Strokes
2
1
1/2
Correction Factor
1/2
1
2
Sampling Time
1 minute per pump stroke
Detecting Limit
0.1 ppm (n=2)
Colour Change
Pale Yellow Pink
Reaction Principle
H2S + HgCl2 HSHgCl + HCl
HCl + Base Chloride
Coefficient of Variation
10% (for 1 to 2 ppm), 5% (for 2 to 6 ppm)
Shelf Life
2 Years
Corrections for temperature & humidity
Unnecessary
Substance
Concentration
Interference
Change colour by itself to
Ammonia
No
No
sulphur dioxide
No
No
Mercaptans
+
Pink
Calibration gas generation
Permeation tube method
TLV-TWA
TLV-STEL
Explosive range
10 ppm
15 ppm
4.0 to 44%
Question/Answer
Do you have a question?
Question (maximum 200 characters):
Email me the answer?
Yes
No
Email address: