Question/Answer

I pressed the arrow keys and I\'m still unable to change the temperature on the controller .

If the contacts under the buttons are no longer responsive you will need to replace the controller.

Where do I put the small red and yellow wires go?

I presume you are referring to the wires that come off of the thermocouple. The yellow wire goes into terminal 1 on the controller and the red wire goes into terminal 2. If you accidentally hook them up in the wrong order your temperature will display negative values.

I HAVE THE 12 PLACE BLOCK #C6002 CAN THE CONTROLLER BE SET FOR 2 DIFFERENT TEMP PROGRAMS?

This controller can only be set to heat and hold at one temperature. There are no ramping options available for it.

Would the sc945-w work with the CEC hot blocks?

This part number will only fit in the HotBlocks that have CECW in the middle of the serial number. The electrical connection will work but the opening to insert the controller is too small on the HotBlocks with CEC in the serial number.

We need a controller for our SC6002 hot block. Which is the correct one to order?

This will be dependent on the serial number of your unit. If the serial number contains the letters \"CECW\" then the correct part # is SC945-W. If the serial number contains the letters \"CEC\" then the correct part # is SC945.

What is the difference between the SC945 and SC945-W controllers?

The only difference in the SC945 and SC945-W controller is a slight variation in the housing dimensions. To determine which one you need you just look at the serial number on the block. If it contains the letters CECW you will need SC945-W. If it is just CEC or any other set of letters you will need SC945.